Troubleshooting
Alarms, Messages and Indications
The Infinity® uses alarms, messages and indications in the display screen to help you diagnose and troubleshoot the most common questions about pump operation.
Troubleshooting - Alarms
LOAD SET
To correct this condition, you would:
- Press RUN/PAUSE to silence alarm and put pump in PAUSE mode.
- Load Infinity® delivery set and close the door.
- Make sure cassette receptacle is clean, especially around sensors.
Troubleshooting - Alarms
LOW BATT
To correct this condition, you would:
- Plug in AC charger.
Troubleshooting - Alarms
NO FLOW IN
To correct this condition, you would:
- Press RUN/PAUSE to silence alarm and put pump in PAUSE mode.
- Check tubing for kinks or blockages.
- Make sure cassette receptacle is clean, especially around sensors.
Troubleshooting - Alarms
NO FLOW OUT
To correct this condition, you would:
- Press RUN/PAUSE to silence alarm and put pump in PAUSE mode.
- Check tubing for kinks or blockages.
- Make sure cassette receptacle is clean, especially around sensors.
- Potential back pressure from patient (check with physician).
Troubleshooting - Alarms
NO FOOD
To correct this condition, you would:
- Press RUN/PAUSE to silence alarm and put pump in PAUSE mode.
- Check the feeding container. Refill and PRIME if empty.
- Check tubing for air bubbles. Disconnect from patient and PRIME to remove air from tubing.
- Make sure cassette receptacle is clean, especially around sensors.
- Make sure Infinity® delivery set is properly loaded and close the door.
- Check delivery set for wear and damage. Replace if necessary.
Troubleshooting - Alarms
PUSH RUN TO FEED
To correct this condition, you would:
- Press RUN/PAUSE to silence alarm and put pump in PAUSE mode.
- Press ON/OFF to turn pump off.
Troubleshooting - Alarms
SHUT DOOR
To correct this condition, you would:
- Press RUN/PAUSE to silence alarm and put pump in PAUSE mode.
- Close pump door and check to make sure the latch is secure.
Troubleshooting - Messages
DOSE DONE
This message indicates a single feeding dose has been delivered.
Troubleshooting - Messages
NEXT DOSE
This message indicates an INTERVAL feeding is underway and the time (in hours and minutes) before the next feeding begins.
Troubleshooting - Indications
CHEK
This message indicates that the programmed RATE, DOSE and FEED INTERVAL are incompatible.
Troubleshooting - Indications
TO STOP LET GO
This message is displayed when pressing and holding the PRIME key to prime the tubing. To stop priming, release the PRIME key.
NOTE: If your pump is equipped with an auto prime option, refer to the Operator?s Manual for details.
Troubleshooting - Indications
Pump on AC Power, battery symbol flashing
This indicates the pump is connected to AC power but the battery is not charging (Contact your equipment provider for service).
Troubleshooting
Complete
You have finished the troubleshooting section. Now you can try your hand at programming the pump.
Go ahead, try it!